medical Class II Updated 2021-04-21

Covidien, LP recalls Covidien Signia Small Diameter Curved Tip Intelligent Reload

Recalled Product

Covidien Signia Small Diameter Curved Tip Intelligent Reload 30 mm Vascular/Thin 8 mm - Short Item Code:SIGSDS30CTVT


Hazard / Issue

Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Numbers: N0A0274UY N0D0037UY N0G0491UY N0K1016UY N0A0897UY N0D0391UY N0G0736UY N0L0077UY N0B0640UY N0E0175UY N0J0510UY N0M0063UY N0B0937UY N0E0802UY N0J0680UY N0M0259UY N0C0144UY N0F1055UY N0J0999UY N9K1094UY N0C0486UY N0G0185UY N0K0584UY N9L0795UY N0C1115UY N0G0250UY N0K0805UY UDI: 10884521741874 20884521741871
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →