medical Class III Updated 2022-07-20

Hamilton Medical AG recalls HAMILTON-H900 Humidifier, Models: 950001, 950004

Recalled Product

HAMILTON-H900 Humidifier, Models: 950001, 950004


Hazard / Issue

When starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not filled into the chamber, the low water level alarm will be delayed and medical staff are unaware that there is no water in the chamber.

Issued by

FDA

Affected States: WI, FL, CA, NV, PA, TX
Lot/Code Info: UDI:07630002801546. All devices with software version 1.10c
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →