medical Class II Updated 2020-03-11

Beckman Coulter Inc recalls Enzymatic Creatinine Reagent (CR-E 2 X 200), REF A60298, For

Recalled Product

Enzymatic Creatinine Reagent (CR-E 2 X 200), REF A60298, For in Vitro Diagnostic Use, Rx Only, UDI: 15099590625733


Hazard / Issue

It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quantities in serum due to acetaminophen overdoese. This could result in a delay in diagnosis and treatment.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All Lot Numbers
View official government recall