medical Class II Updated 2020-03-11

Beckman Coulter Inc recalls Triglycerides Reagent (TG 2 x 300), REF: 445850, For In Vitr

Recalled Product

Triglycerides Reagent (TG 2 x 300), REF: 445850, For In Vitro Diagnostic Use, Rx Only, UDI: 15099590233228


Hazard / Issue

It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quantities in serum due to acetaminophen overdoese. This could result in a delay in diagnosis and treatment.

Issued by

FDA

Distribution: Nationwide (United States)