medical Class II Updated 2020-03-11

RESPIRATORY THERAPEUTICS GROUP LLC recalls RespVent Closed Suction System T-Piece Tracheostomy Adult, D

Recalled Product

RespVent Closed Suction System T-Piece Tracheostomy Adult, Diameter: 4.6mm (14F), Length: 30.5cm (12in.), MDI Adapter, Cont. 10 Closed Suction Catheter, Sterile R, RxOnly, Reorder Number: RTG-02300, UDI: 00850108006035 - Product Usage: To be used to aspirate liquids or semi solids from a patient s upper airway.


Hazard / Issue

It has been determined that some units of catheters were not tightly connected to the thumb port. This may result in the catheter disconnecting from the thumb port which could reduce the suction performance and make it difficult to remove from the patient's trachea.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot KS1901013: 20019/03/31 - 06/192019 Lot KS1902002: 2019/06/21 - 07/31/2019
View official government recall