medical Class II Updated 2022-07-27

Aesculap Implant Systems LLC recalls AESCULAP S4 Cervical Cross Connector, L26mm, S4C CROSS CONNE

Recalled Product

AESCULAP S4 Cervical Cross Connector, L26mm, S4C CROSS CONNECTOR FIXED 26MM, product code SW114T. Used in spinal fusion surgery.


Hazard / Issue

Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement of the connectors on the rods and/or microparticle abrasion.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 04038653362318
View official government recall