medical Class II Updated 2018-05-09

Siemens Medical Solutions USA, Inc recalls SOMATOM Spirit Intended to produce cross-sections images

Recalled Product

SOMATOM Spirit Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.


Hazard / Issue

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Issued by

FDA

Distribution: USA (nationally) Distribution.
Lot/Code Info: (Not supplied)
View official government recall