medical Class II Updated 2020-03-18

Cardinal Health 200, LLC recalls Monoject Hypodermic Safety Needle 25G x 5/8" Item Code 1182

Recalled Product

Monoject Hypodermic Safety Needle 25G x 5/8" Item Code 1182558 - Product Usage: used for general purpose injection and aspiration of fluids from vials, ampules and parts of the body below the surface of the skin.


Hazard / Issue

Non-sterile product was shipped to customers. Product package is labeled as both "STERILE" and "NON-STERILR" ; the product did not undergo sterilization process and is non-sterile. The usage of a needle that is not sterile could result in infection. Firm is not aware of any reports of patient harm.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot No. 15063004, Item Code 1182558, Expires 5/31/2020