medical Class II Updated 2018-04-25

Cook Inc. recalls Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN

Recalled Product

Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275. Intended for the aspiration of body fluids and cells, in particular oocyte aspiration.


Hazard / Issue

Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.

Issued by

FDA

Distribution: Worldwide Distribution -- USA, Argentina, Bolivia, Brazil, Canada, Colombia, Guatemala, Mexico, Panama, Peru, Uruguay, and Venezuela.
Lot/Code Info: All lots