medical Class II Updated 2020-03-18

Response Biomedical Corp. recalls Randox Laboratories Ltd. Catalogue Number CQ5052, Lot Number

Recalled Product

Randox Laboratories Ltd. Catalogue Number CQ5052, Lot Number 4244CK (relabeled as Response Biomedical Catalogue Number C2003-2, Lot #070318) IFU: "RAMP Cardiac Controls *** CAT. No.: C2003-2 *** Size: 3 x 3 ML". Product Usage: use in the quality control of cardiac marker assays run on the RAMP¿ platform.


Hazard / Issue

The form received three complaints related to the user obtaining a result that fell below the target range for CK-MB provided in the RAMP Cardiac Controls Instructions for Use. Via in-house monitoring, determined that liquid Cardiac Control has seen a decrease in CK-MB Levels of up to 57% signal loss within 11 months. Shelf-life of product is 24 months.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Randox Laboratories Ltd. Catalogue Number CQ5052, Lot Number 4244CK (relabeled as Response Biomedical Catalogue Number C2003-2, Lot #070318). UDI Code: (01) 1 0627966 00508 4 (10) 070318 (17)300919
View official government recall