medical Class II Updated 2020-03-18

RAYSEARCH LABORATORIES AB recalls RayStation/RayPlan 8B, Build Number 8.1.0.47, stand-alone so

Recalled Product

RayStation/RayPlan 8B, Build Number 8.1.0.47, stand-alone software treatment planning system


Hazard / Issue

There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.

Issued by

FDA

Affected States: TN
Lot/Code Info: UDI 07350002010129
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →