medical Class II Updated 2020-03-18

RAYSEARCH LABORATORIES AB recalls RayStation/RayPlan 8B Service Pack 1, Build Number 8.1.1.8,

Recalled Product

RayStation/RayPlan 8B Service Pack 1, Build Number 8.1.1.8, stand-alone software treatment planning system


Hazard / Issue

There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.

Issued by

FDA

Affected States: TN
Lot/Code Info: UDI 07350002010204
View official government recall

Looking for safe alternatives?