medical Class II Updated 2025-04-02

Checkpoint Surgical Inc recalls Brand Name: Checkpoint Guardian Intraoperative Lead Product

Recalled Product

Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead, Small Model/Catalog Number: REF# 9524 (Small) Product Description: The Checkpoint Guardian Intraoperative Leads are single-patient disposable accessories for providing stimulus to a targeted nerve. The accessories are intended for use with any device from the Checkpoint Stimulator/Locator Family. The lead electrode is intended to wrap around nerves that have been surgically exposed, allowing the surgeons to provide hands free stimulus to a targeted nerve.


Hazard / Issue

Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed metal. There is a risk of electrical current leakage, which may lead to inconsistent or confusing muscle responses, potentially compromising the effectiveness of the intended use as the surgeon could take an incorrect clinical action.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI B13095240 Small Lot 4318 Expires 01Apr2026 Lot 4325 Expires 01Apr2026 Lot 4368 Expires 01Dec2026