medical Class II Updated 2026-03-04

Medline Industries, LP recalls Convenience kits containing Tego Connectors Medline DIALY

Recalled Product

Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167


Hazard / Issue

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI (ea) 10195327015954 UDI-DI (case) 40195327015955 Kit lots 24IME953 25CMA708
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →