medical Class II Updated 2026-03-04

Agiliti Health - Ellis recalls Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, compo

Recalled Product

Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure


Hazard / Issue

Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI 0084569904914; All serial numbers with a manufacturing date prior to 12/04/2025