medical Class II Updated 2025-04-09

Stryker Corporation recalls Nasopore Ex Firm 4cm fragmentable nasal dressing, Catalog Nu

Recalled Product

Nasopore Ex Firm 4cm fragmentable nasal dressing, Catalog Number 5400-030-004 and Nasopore Forte plus 4cm Catalog Number 5400-030-004ITL


Hazard / Issue

There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: 5400-030-004 UDI-DI: 07613154340835, Lot Numbers: 2024052333 2023012514 2023050514 2023080912 2024020214; 5400-030-004ITL UDI-DI: 07613327359442, Lot Numbers: 2023071933 2023102714 2024012314
View official government recall