medical Class II Updated 2026-03-11

Staar Surgical AG recalls EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: V

Recalled Product

EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5


Hazard / Issue

Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.

Issued by

FDA

Distribution: International distribution to the countries of India, Iran, Korea, Saudi Arabia.
Lot/Code Info: UDI-DI: 00840311304363. Expiration Date: 2027-11-30 Serial Numbers: T2286964, T2286971, T2286973, T2286979, T2286980, T2286981, T2286988