medical Class I Updated 2018-05-16

Draeger Medical, Inc. recalls Fabius MRI, anesthesia machine, catalog no. 8607300 Produ

Recalled Product

Fabius MRI, anesthesia machine, catalog no. 8607300 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). This device can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less.


Hazard / Issue

The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial no: ASKM-0020 ASKM-0011 ASKL-0005 ASKL-0007 ASKL-0010 ASKM-0031 ASKM-0032 ASKM-0023
View official government recall