medical Class I Updated 2018-05-16

Draeger Medical, Inc. recalls Fabius Tiro M, anesthesia machine, catalog no. 8608400 Pr

Recalled Product

Fabius Tiro M, anesthesia machine, catalog no. 8608400 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2).


Hazard / Issue

The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial no. ASKK-0001
View official government recall