medical Class II Updated 2026-03-11

Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urine Albumin (UAlb). Material Number: 11537225

Recalled Product

Atellica CH Urine Albumin (UAlb). Material Number: 11537225


Hazard / Issue

Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Material Number: 11537225. UDI-DI: 00630414611099. Lot Numbers: All lot numbers.
View official government recall