medical Class II Updated 2020-03-25

Philips North America, LLC recalls Cardio Vascular-Allura Centron, system code 722400

Recalled Product

Cardio Vascular-Allura Centron, system code 722400


Hazard / Issue

A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Not distributed in US. Serial numbers 132 133 49 66 71 135 31 36 79 69 137 149 147 109 136 108 103 93 53 127 129 129 55 74 2 57 106 56 86 68 157 78 64 63 23 51 100 97 102 113 90 44 30 101 77 83 67 151 22 70 96 116 131 35 82 9 91 43 61 161 28 34 27 19 144 92 168 172 173 122 197 139 6 76 176 16 130 60 80 184 112 81 89 42 110 155 33