medical Class II Updated 2019-06-05

Maquet Cardiovascular, LLC recalls Maquet Getinge-BEQ-TOP 5103#CPS Tubing Pack Material: 70104

Recalled Product

Maquet Getinge-BEQ-TOP 5103#CPS Tubing Pack Material: 701048690


Hazard / Issue

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Batch Numbers: 3000044416 3000049077 3000053102 3000054957 3000056257 3000061306 3000070701 3000072685
View official government recall

Looking for safe alternatives?