medical Class I Updated 2025-04-16

Qapel Medical Inc. recalls Hippo 072 Aspiration System including Cheetah Delivery Tool

Recalled Product

Hippo 072 Aspiration System including Cheetah Delivery Tool and Aspiration Tubing, REF: APT6072-132. Hippo 072 Aspiration System including Cheetah Delivery Tool, REF: AP6072-132. 072 Aspiration Tubing, REF: APT-95


Hazard / Issue

Aspiration catheter distal tip features and characteristics may not be in the scope of FDA clearance; tip detachment, vessel rupture, and vasospasm could occur.

Issued by

FDA

Affected States: CA, IL, NE, TN, TX, NJ, NC, SD, FL, MI, PA, GA, AL, IN, NY, WI, AZ, MN, LA, OH, OK, SC, NV, ND, ID, MO, CT, WA, MA
Lot/Code Info: REF/UDI-DI/Lot(Expiration): APT6072-132/00857545008127/FG241008C-03(04/07/2025), FG240916C-04(03/17/2025), FG240905C-04(03/06/2025). AP6072-132/00857545008097/FG241206A-03(06/08/2025), FG240917A-01(03/17/2025). APT-95/00857545008103/FG241206A-04(06/08/25)
View official government recall