Maquet Cardiovascular, LLC recalls Maquet Getinge-BEQ-TOP 15705 Material: 701051122

Recalled Product

Maquet Getinge-BEQ-TOP 15705 Material: 701051122


Hazard / Issue

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Batch Numbers: 3000044566 3000046394 3000049202 3000054550 3000058043 3000060239 3000066276 3000069368 3000071833 3000074339 3000080701 3000081598