medical Class III Updated 2018-05-02

Diagnostic Hybrids, Inc. recalls Quidel MicroVue Intact PTH EIA, Model 8044. An enzyme immun

Recalled Product

Quidel MicroVue Intact PTH EIA, Model 8044. An enzyme immunoassay for the quantification of Intact Parathyroid Hormone (PTH) in serum.


Hazard / Issue

There may be a loss of separation between Calibrators A and B which may impact the assay calibration at the lower end of the assay.

Issued by

FDA

Distribution: The product was distributed to the following countries: Singapore, Switzerland
Lot/Code Info: Lot 068259