medical Class II Updated 2018-05-02

Sofradim Production recalls Versatex Monofilament Mesh 50 x 50cm, Product No. VTX5050M

Recalled Product

Versatex Monofilament Mesh 50 x 50cm, Product No. VTX5050M The device is made out of a macroporous three-dimensional monofilament polyester textile. Its intended use is for the repair of abdominal wall hernias or other fascial deficiencies that require the addition of a reinforcing material.


Hazard / Issue

There have been patient reports of abdominal hernia recurrence following hernia repair. The recalling firm has added a statement to the device's Instructions for Use to make surgeons aware of risks in using the device in Transversus Abdominis Muscle Release (TAR) procedures.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI - 10884521532298 Lots SPC1091X SPJ0706X SPJ1016X SQA0990X SQI0302X SQJ0443X SQK0319X SRA2188X SRI0548X
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →