medical Class II Updated 2021-05-12

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Implantable cardiac resynchronization therapy pacemakers (CR

Recalled Product

Implantable cardiac resynchronization therapy pacemakers (CRT-P) - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. Percepta family: Percepta CRT-P MRI SureScan, Percepta Quad CRT-P MRI SureScan. 2. Serena family: Serena CRT-P MRI SureScan, Serena Quad CRT-P MRI SureScan. 3. Solara family: Solara CRT-P MRI SureScan, Solara Quad CRT-P MRI SureScan.


Hazard / Issue

A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

Issued by

FDA

Distribution: Worldwide distribution.
Lot/Code Info: all codes