medical Class II Updated 2021-05-12

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls CareLink SmartSync Device Manager - Product Usage: used to t

Recalled Product

CareLink SmartSync Device Manager - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure, Model Number 24970A


Hazard / Issue

A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

Issued by

FDA

Distribution: Worldwide distribution.
Lot/Code Info: all codes
View official government recall