medical Class II Updated 2019-06-05

Maquet Cardiovascular, LLC recalls Maquet Getinge-BEQ-TOP 25001 MONTEFIORE Material: 701053906

Recalled Product

Maquet Getinge-BEQ-TOP 25001 MONTEFIORE Material: 701053906


Hazard / Issue

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Batch Numbers: 3000054551 3000065085
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →