medical Class II Updated 2021-05-12

Greiner Bio-One North America, Inc. recalls VACUETTE¿ Blood Collection tube 9NC Coagulation sodium citra

Recalled Product

VACUETTE¿ Blood Collection tube 9NC Coagulation sodium citrate 3.2% 13x75 blue cap-white ring, sandwich tube, non-ridged - Product Usage: indicated for use in collection and transport of venous whole blood for coagulation testing.


Hazard / Issue

VACUETTE¿ Blood Collection tube 9NC Coagulation sodium citrate 3.2% may show insufficient draw volume of the tubes, which could produce erroneous results for prothrombin time, activated partial thromboplastin time, and fibrinogen assays.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Item Number: 454322; Lot # A21023B5; UDI: (Case) 39120017574485 and (Rack) 29120017574488; Expiration Date: 02/06/2022 Item Number: 454332; Lot # A210338Q; UDI: (Case) 39120017574607 and (Rack) 29120017574600; Expiration Date: 03/01/2022 Item Number: 454334; Lot # A210233V; UDI: (Case) 39120017574539 and (Rack) 29120017574532; Expiration Date: 02/01/2022 Item Number: 454334; Lot # A21023XW; UDI: (Case) 39120017574539 and (Rack) 29120017574532; Expiration Date: 02/13/2022