medical Class II Updated 2019-06-05

Maquet Cardiovascular, LLC recalls Maquet Getinge-BEQ-TOP 37200 CIRCULATORY CIRC Material: 7

Recalled Product

Maquet Getinge-BEQ-TOP 37200 CIRCULATORY CIRC Material: 701055388


Hazard / Issue

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Batch Numbers: 3000044291 3000046812 3000048395 3000051270 3000054144 3000055979 3000058735 3000061304 3000063708 3000066691 3000070702 3000072307 3000075122