medical Class II Updated 2024-04-24

Smiths Medical ASD Inc. recalls Smiths Medical Medfusion Guide Barrel Clamp, Item number G60

Recalled Product

Smiths Medical Medfusion Guide Barrel Clamp, Item number G6000716


Hazard / Issue

Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.

Issued by

FDA

Distribution: Worldwide distribution.
Lot/Code Info: Lot Numbers: P0407365, P0486670, P0561740
View official government recall