medical Class II Updated 2019-06-05

Maquet Cardiovascular, LLC recalls Maquet Getinge-BEQ-TOP 38900 ECC PACK Material: 701056368

Recalled Product

Maquet Getinge-BEQ-TOP 38900 ECC PACK Material: 701056368


Hazard / Issue

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Batch Numbers: 3000041396 3000044053 3000047317 3000050887 3000058131 3000060487 3000060723 3000062458 3000064724 3000071815 3000071817
View official government recall