Siemens Medical Solutions USA, Inc recalls AXIOM Sensis or Sensis Vibe Combo systems as follows: Syste
Recalled Product
AXIOM Sensis or Sensis Vibe Combo systems as follows: System Material # AXIOM Sensis, Combo 64 6623974; AXIOM Sensis, Hemo LOW 6634633; AXIOM Sensis, Combo 32 6634641; AXIOM Sensis, EP 129 6634658; Sensis 10764561; Sensis Vibe Combo 11007642; Sensis recording system is intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology.
Hazard / Issue
The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility of mounting the Signal Input Box on an accessory rail (e.g., at the patient table), In the event the mounting position deviates from the default position, the following are possible: Patient touching the Signal Input Box may result in a touch leakage current to the patient; Spillage of liquid entering the Signal Input Box