medical Class II Updated 2019-06-05

Maquet Cardiovascular, LLC recalls Maquet Getinge-BEQ-TOP 24105 ECC PACK Material:701064757

Recalled Product

Maquet Getinge-BEQ-TOP 24105 ECC PACK Material:701064757


Hazard / Issue

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Batch Numbers: 3000051271 3000057632 3000062460 3000066202 3000069076 3000073425 3000076761 3000080703