medical Class II Updated 2019-06-05

Maquet Cardiovascular, LLC recalls Maquet Getinge-BEQ-TOP 31203 CUSTOM PACK NEO Material: 70

Recalled Product

Maquet Getinge-BEQ-TOP 31203 CUSTOM PACK NEO Material: 701065510


Hazard / Issue

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Batch Numbers: 3000050481 3000052617 3000054413 3000054961 3000055976