medical MODERATE Updated 2026-03-25

Baxter Healthcare Corporation recalls Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spo

See all recalls from Baxter Healthcare Corporation →

Recalled Product

Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software Version: Software version 3.2.0.1 Product Description: Welch Allyn Spot Vision Screener VS100, FDA product code name: Refractometer, ophthalmic One unit per package

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Code: " Product Code: HKO " Version or Model: 901029 " Catalog Number: VS100 " Company Name: WELCH ALLYN, INC. " Primary DI Number: 00732094214154 " Packaging DI Number: 00732094212228 " Device Description: VS100 UDI BASE UNIT " Software version: 3.2.0.1 " Serial Numbers: Refer to attached Affected Product Table

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for Brand Name: Welch Allyn, →
View official government recall