medical Class II Updated 2020-04-08

Stryker Communications recalls Berchtold Chromophare Ceiling Mounted Surgical Light System,

Recalled Product

Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000


Hazard / Issue

The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Number Range: 091316 70016 through 083117-73252