medical Class II Updated 2018-05-09

Philips Medical Systems Nederlands recalls MultiDiagnost Eleva 708036 As a multifunctional universal

Recalled Product

MultiDiagnost Eleva 708036 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.


Hazard / Issue

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: System Serial # 18 42 46 41 36 35 37 47 27 43 10 5 24 29 34 49 26 13 15 25 50 12 17 48 32 30 31 14 40 19 9 16 45 8 4 21 11