medical Class II Updated 2018-05-09

Philips Medical Systems Nederlands recalls MultiDiagnostic-Eleva with Flat Detector 708035 As a mult

Recalled Product

MultiDiagnostic-Eleva with Flat Detector 708035 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.


Hazard / Issue

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Pending
View official government recall