medical Class II Updated 2022-08-31

Mckesson Medical-Surgical Inc. Corporate Office recalls a. BECTON DICKINSON BD Vacutainer SST TUBE, BLD COL CLOT/SEP

Recalled Product

a. BECTON DICKINSON BD Vacutainer SST TUBE, BLD COL CLOT/SEP GLD 3.5ML PLAS VACUTAINER (100/BX), Model Number: 367983. b. BECTON DICKINSON BD Vacutainer SST TUBE, BLD COL CLOT/SEP RED/GRY8.5ML PLAS VACUTAINER (100/BX Model Number: 367988. c. BECTON DICKINSON BD Vacutainer SST TUBE, BLD COL CLOT/SEP GLD 5MLPLAS VACUTAINER (100/BX) Model Number: 367986. d. BECTON DICKINSON BD Vacutainer SST TUBE, BLD COL CLOT/SEP RED/GRY7.5ML PLAS VACUTAINER (100/BX) Model Number: 367987. e. BECTON DICKINSON BD Vacutainer SST TUBE, BLD COL CLOT/SEP RED/GRY3.5ML PLAS VACUTAINER (100/BX Model Number: 367981. f. BECTON DICKINSON BD Vacutainer TUBE, BLD COL PLAIN RED 10ML GLS VACUTAINER (100/BX 10BX/CS) Model Number: 366430. g. BECTON DICKINSON BD Microtainer SST - Amber TUBE, CAPILLARY CLOT/SEP GLD 400/600UL PLAS MICROT (50/BX) Model Number: 365978. h. BECTON DICKINSON BD Vacutainer CPT NC: 1.0mL FICOLL: 2.0mL TUBE, BLD CPT SOD CIT 8ML (60/CS) Model Number: BD 362761. i. BECTON DICKINSON BD Vacutainer Rapid Serum Tube (RST) Thrombin TUBE, BLD COL CLOT ORG 5ML PETVACUTAINER (100/BX 10BX/CS) Model Number: 368774. j. BECTON DICKINSON BD Vacutainer Buff. Na Citrate 0.109M, 3.2% TUBE, BLD COL SODCIT LTBLU 2.7ML PLAS VACUTAINER (100/BX) Model Number: 363083.


Hazard / Issue

Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: GTIN: a. 00382903679836; 50382903679831. b. 00382903679881; 50382903679886. c. 00382903679867; 50382903679862. d. 00382903679874; 50382903679879, e. 00382903679812; 50382903679817. f. 00382903664306; 30382903664307; 50382903664301. g. 00382903659784; 50382903659789. h. 00382903627615; 50382903627610. i. 00382903687749; 50382903687744. j. 00382903630837; 50382903630832.
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →