medical Class II Updated 2020-04-08

Respironics California, LLC recalls Philips V60 Ventilators with Power Management PCBA part numb

Recalled Product

Philips V60 Ventilators with Power Management PCBA part number 1055906


Hazard / Issue

Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia

Issued by

FDA

Affected States: CO, PA, AZ, OH, MT, AR, FL, NC, KS, IL, CA, HI, MD, SC, WY, OR, NY, NE, OK, MN, UT, ND, LA, IN, AL, AK, WV, GA, WI, MI, VA, NJ, TX, WA, SD, KY, TN, MS, MO, DE, MA, VT, CT, IA, NV, ID, ME, NM, DC, NH, RI
Lot/Code Info: Serial Numbers: 100002908 to 100027093; 100027096 to 100085389; 100085399 to 100109746; 100109868 to 201003390; 201003392 to 201010952
View official government recall