medical Class II Updated 2021-06-02

Philips Respironics, Inc. recalls BiPAP V30 Auto, US market: Continuous ventilator, non-life-s

Recalled Product

BiPAP V30 Auto, US market: Continuous ventilator, non-life-supporting. Material Number: 1111178


Hazard / Issue

Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Number: N3059246815B4, N305925228752, N30592449378D, N3059251913E9, N30592514C80C, N30592513BCB3, N30592515D985, N305924084164, N305925119FA1, N30592516EB1E, N305925180260, N30592457C72B, N306372301F1F, N306372551C62, N30637223345C, N30592416B1C2, N30592524E264, N30592520A440, N3059252396DB, N30592517FA97, N3059241492D0, N30592512AD3A.