medical Class II Updated 2021-06-02

Philips Respironics, Inc. recalls BiPAP A40, China: Continuous ventilator, non-life-supportin

Recalled Product

BiPAP A40, China: Continuous ventilator, non-life-supporting. Material Number - CN1111169


Hazard / Issue

Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Number: V30446637ACAE, V30446653BE5A, V304441360854, V304457587F98, V304468408502, V304441766F34, V3044555732D7, V30446846E034, V30444034719A, V3044642346EA.