medical Class II Updated 2021-06-02

Philips Respironics, Inc. recalls BiPAP A40 System Silver, JP: Continuous ventilator, non-life

Recalled Product

BiPAP A40 System Silver, JP: Continuous ventilator, non-life-supporting. Material Number: R1111177


Hazard / Issue

Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Number: V30562261B40E
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →