Philips Respironics, Inc. recalls BiPAP A30, Brazil: Continuous ventilator, non-life-supportin
Recalled Product
BiPAP A30, Brazil: Continuous ventilator, non-life-supporting. Material Number: BR1111143
Hazard / Issue
Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.