medical Class II Updated 2018-05-16

Medtronic Sofamor Danek USA Inc recalls Medtronic Prestige LP(TM) Cervical Disc System with Streamli

Recalled Product

Medtronic Prestige LP(TM) Cervical Disc System with Streamlined Instruments, DRILL GUIDES: a) REF 6975205, 5mm b) REF 6975206, 6mm c) REF 6975207, 7mm Product Usage: The PRESTIGE LP reusable instruments are intended for the surgical implantation of the PRESTIGE LP Cervical Disc. Medtronic reusable instruments must be sterilized prior to initial use and must be cleaned and sterilized prior to each re-use.


Hazard / Issue

The firm received complaints of drill bits breaking during use with the Prestige LP(TM) Cervical Disc System. Subsequent investigation demonstrated the interaction of the drill guide and the drill bit may lead to drill bit breakage.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: a) REF 6975205, 5mm; UDI 00643169752979 Lot Number ID17K018 b) REF 6975206, 6mm; UDI 00643169752962 Lot Number ID17J009 c) REF 6975207, 7mm; UDI 00643169752955 Lot Numbers: ID17K007, ID17L035
View official government recall

Looking for safe alternatives?