medical Class II Updated 2018-05-16

Deerfield Imaging, Inc. recalls IMRIS IMRI 1.5T AND 3T S System Product Usage: The IMRIS

Recalled Product

IMRIS IMRI 1.5T AND 3T S System Product Usage: The IMRIS intraoperative MRI systems are traditional MRI systems that have been suspended on an overhead rail system. and is designated to operative inside an RF shielded room to facilitate intraoperative and multi room use.


Hazard / Issue

The DC power to the IFOSS collision detector control panel at your facility is being adversely affected by the magnetic field. Our investigation shows the magnetic field is causing the IFOSS control panel to malfunction.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers: 10111-00, 10073-01, 10043-01, 10097-01, 10006-11, 10076-21, 10081-01, 10089-01, 10101-01, 10093-01