medical Class II Updated 2025-04-30

Zimmer, Inc. recalls Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +

Recalled Product

Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +0 Neck, Model Number 802602202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented


Hazard / Issue

The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: GTIN 00887868376320, All lots distributed with IFU 87-6204-051-99 Rev D or prior