medical Class II Updated 2019-06-05

Maquet Cardiovascular, LLC recalls Maquet Getinge: (1) Material: 701053941R01 BEQ-TOP 9005 3

Recalled Product

Maquet Getinge: (1) Material: 701053941R01 BEQ-TOP 9005 3/8 Custom Pk w/RF32 & HMOD (2) Material: 701053941R02- BEQ-TOP 9005 3/8 Custom Pk w RF32 & HMOD Material:


Hazard / Issue

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Batch Numbers: (1) 3000051652 3000054808 3000058733 3000062473 3000071201 3000073170 3000075123 (2) 3000076148 3000081600
View official government recall